本文主要为广大网友提供“心电系统的设计”,希望对需要心电系统的设计网友有所帮助,学习一下!
论文编号:DQ306 论文字数:16189,页数:38
摘 要
本文主要对心电检测系统的设计做了详细的介绍。整个系统包括心电信号的采集、A/D转换、液晶显示三部分。基于对心电信号阻抗高、幅度小和频率低等特点的考虑,我们采用二级放大、低通滤波器对其进行放大和频带压缩;对于电路中严重的共模干扰,我们采用右腿驱动电路、电缆驱动技术来进行处理。经过以上几个关键的处理后,再送入单片机进行A/D转换和液晶显示。整个系统以AD620作为放大电路核心器件;以MSP430F449作为微处理控制芯片;采用RT12864M液晶模块显示心电图。
关键字:心电图、二级放大、右腿驱动、MSP430
Abstract
The thesis mainly introduced the design of a system, which can be used to detect and depict the electrocardiogram (ECG). The system comprises three pacts: the detecting of ECG, A/D converter and LCD display. The source of electrocardiogram is in the performance of high resistance, small magnitude and low frequency, due to which the amplificatory circuit, low-pass filter are used to amplify the signal and attenuate the bandwidth; Also the Driven-Right-Leg Circuit and the Driven Cable technique are used to attenuate the common-mode disturbance. After the process of what mentioned above, the microprocessor can convert and display the ECG. In this design, AD620 is the core device of the amplification; MSP430 is used as microprocessor; RT12864M is used for the display of ECG.
Key words: Electrocardiogram Double-stage amplify Driven-Right-Leg Circuit MSP430
目 录
摘 要 I
Abstract II
目 录 III
1 绪 论 1
1.1引言 1
1.2心电信号简介 1
1.2.1心电及其记录方法的早期研究 1
1.2.2心电的产生机理 3
1.2.3心电信号的电特点 5
1.2.4心电信号的临床应用 5
2总体设计 7
2.1系统基本设计思想 7
2.2系统的工作原理及组成 7
2.2.1生物电测量电极 7
2.2.2体表心电图导联 7
2.2.3 前置放大器的设计要求 8
2.2.4系统软件功能 9
3系统模拟部分 11
3.1前置放大电路 11
3.2 50HZ工频抑制电路 13
3.2.1工频干扰原理分析 13
3.2.2解决途径之―右腿驱动电路 14
3.2.3解决途径之二-电缆屏蔽驱动电路 16
3.3滤波电路的设计 17
3.4二级放大电路 20
4系统数字部分 22
4.1单片机的选择 22
4.2 A/D转换部分 23
4.3 LCD显示部分 24
4.3.1液晶与显示 25
4.3.2 LCD与MSP430的硬件接口 26
4.3.3 MSP430F449资源使用 27
5测试部分 29
6总结和讨论 30
致 谢 31
参考文献 32
附 录 34
附录1 原理图: 34
附录2 元件清单: 35